What is "Fluorescence"?

The TEII Dissolved Oxygen Fluoroprobe together with the Fluorometer controller is designed to provide long-term monitoring and control of appropriate devices based on the saturation level of oxygen in gaseous or aqueous solutions.  The standard unit comes with a contact for aerator control.

Dissolved Oxygen MeterDissolved Oxygen Sensor

Dependable - The TEII-DO Dissolved Oxygen Meter can be used for long-term monitoring of oxygen levels without the necessity of changing the membrane or electrolyte fluids. The TEII-DO unit uses optics instead of the other methods which make the TEII-DO more reliable.

Easy To Use - Does not require repetitive calibration and/or compensation for membrane effects which are an extremely large source of error and drift for any probe using membrane/electrolyte and dissimilar metal anodes and cathodes (galvanic and "Clark probe").

Reliable - The maintenance requirement for continued long-term utilization is to wipe off the sensor when necessary.

Versatile - Can be configured to control aeration of wastewater operations, aqua culture applications and process control.

Economical - The Fluoroprobe together with the Fluorometer controller can be provided at a cost which normally will be recovered by cost savings of power consumption.

Aeration Control - TEII-DO1 has a standard contact closure for aeration operation based on a high/low setting on the controller.

Available Options:

  • 4-20ma Current Analog or 0-10V Voltage Output of the Real DO Value
  • Remote control via Analog signal of the High/Low contact closure values
  • Integrated Remote Control and Monitoring of DO and Aeration

 

Model Number

Description

Code Number

Price

TEII-DO1

Self Contained Enclosure

A Contact TEII
TEII-DO2 Panel Mount DO A Contact TEII
 
* TEII-DO1 and TEII-DO2 are equipped standard with a 12 ft. cable length to the sensor.


Equipment Specifications

Application Temperature Range: Frog On A Stick
Instrument -30oC to + 50oC
Sensor Material -85oF to +400oF
Operation Range:
Aqueous Solutions    0ppm to Supersaturation

Response Time:

0

Sensor Thickness

.001"   .005"  .010"  .020"
Aqueous Solutions
*(see footnote below)

30 sec

1 min   2.5 min  10 min
Gaseous Phases 1 10 30 80 sec
Accuracy:
Aqueous Solutions-Range    0 to 3ppm 4 to 7ppm 8 to 10ppm 11ppm to Supersaturation
Accuracy .05ppm .075ppm .1 to.15ppm .2ppm
Gaseous Phases - Range   0% to 25% 26% to 75% 76% to 100%
Accuracy 1% 1.5% 2%
Instrument Measurement Steady State (Amplitude)/Lifetime
Temperature Compensation Yes
Ambient Excitation Compensation Yes
Sensor Lifetime (Glass/Bonded/Non-Bonded) Based solely upon the application environment of Silicone Rubber
NOTE 1: Response time based upon approximately 1 ft/minute flow across sensor- Response time will decrease with increased flow, however, FLOW IS NOT REQUIRED FOR MEASUREMENTS

What is "Fluorescence"?